Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
205User Manual 1173.9240.02 ─ 17
Manual operation: See "Full Scale Level" on page 70
INPut:DIQ:RANGe:COUPling <State>
If enabled, the reference level for digital input is adjusted to the full scale level automat-
ically if the full scale level changes.
This command is only available if the optional Digital Baseband Interface (R&S FSW-
B17) is installed.
Parameters:
<State> ON | OFF
*RST: OFF
Manual operation: See "Adjust Reference Level to Full Scale Level" on page 70
INPut:
DIQ:RANGe[:UPPer] <Level>
Defines or queries the "Full Scale Level", i.e. the level that corresponds to an I/Q sam-
ple with the magnitude "1".
This command is only available if the optional Digital Baseband Interface (R&S FSW-
B17) is installed.
Parameters:
<Level> <numeric value>
Range: 1 μV to 7.071 V
*RST: 1 V
Manual operation: See "Full Scale Level" on page 70
INPut:DIQ:RANGe[:UPPer]:UNIT <Unit>
Defines the unit of the full scale level (see "Full Scale Level" on page 70). The availa-
bility of units depends on the measurement application you are using.
This command is only available if the optional Digital Baseband Interface (R&S FSW-
B17) is installed.
Parameters:
<Level> VOLT | DBM | DBPW | WATT | DBMV | DBUV | DBUA | AMPere
*RST: Volt
Manual operation: See "Full Scale Level" on page 70
INPut:DIQ:SRATe <SampleRate>
This command specifies or queries the sample rate of the input signal from the Digital
Baseband Interface (R&S FSW-B17, see "Input Sample Rate" on page 69).
Configuring the Measurement