Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
210User Manual 1173.9240.02 ─ 17
Query parameters:
<Type> The power sensor type, e.g. "NRP-Z81".
<Interface> Currently not evaluated
Return values:
<Placeholder> Currently not used
<Type> Detected power sensor type, e.g. "NRP-Z81".
<Interface> Interface the power sensor is connected to; always "USB"
<SerialNo> Serial number of the power sensor assigned to the specified
index
Example:
SYST:COMM:RDEV:PMET2:DEF '','NRP-Z81','',
'123456'
Assigns the power sensor with the serial number '123456' to the
configuration "Power Sensor 2".
SYST:COMM:RDEV:PMET2:DEF?
Queries the sensor assigned to "Power Sensor 2".
Result:
'','NRP-Z81','USB','123456'
The NRP-Z81 power sensor with the serial number '123456' is
assigned to the "Power Sensor 2".
Manual operation: See "Select" on page 82
Configuring Power Sensor Measurements
CALibration:PMETer<p>:ZERO:AUTO ONCE...................................................................210
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]........................................................... 211
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE.......................................211
CALCulate<n>:PMETer<p>:RELative:STATe....................................................................212
FETCh:PMETer<p>?......................................................................................................212
READ:PMETer<p>?....................................................................................................... 212
[SENSe:]PMETer<p>:DCYCle[:STATe]............................................................................ 212
[SENSe:]PMETer<p>:DCYCle:VALue.............................................................................. 213
[SENSe:]PMETer<p>:FREQuency...................................................................................213
[SENSe:]PMETer<p>:FREQuency:LINK...........................................................................213
[SENSe:]PMETer<p>:MTIMe.......................................................................................... 214
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt................................................................214
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]...............................................................215
[SENSe:]PMETer<p>:ROFFset[:STATe]...........................................................................215
[SENSe:]PMETer<p>[:STATe].........................................................................................215
[SENSe:]PMETer<p>:UPDate[:STATe]............................................................................ 216
UNIT<n>:PMETer<p>:POWer......................................................................................... 216
UNIT<n>:PMETer<p>:POWer:RATio............................................................................... 216
CALibration:PMETer<p>:ZERO:AUTO ONCE
This commands starts to zero the power sensor.
Configuring the Measurement