Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
232User Manual 1173.9240.02 ─ 17
Parameters:
<Frequency> The allowed range and f
max
is specified in the data sheet.
UP
Increases the center frequency by the step defined using the
[SENSe:]FREQuency:CENTer:STEP command.
DOWN
Decreases the center frequency by the step defined using the
[SENSe:]FREQuency:CENTer:STEP command.
*RST: fmax/2
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Usage: SCPI confirmed
Manual operation: See "Center Frequency" on page 72
See "Center frequency" on page 93
[SENSe:]FREQuency:CENTer:STEP <StepSize>
This command defines the center frequency step size.
You can increase or decrease the center frequency quickly in fixed steps using the
SENS:FREQ UP AND SENS:FREQ DOWN commands, see [SENSe:]FREQuency:
CENTer on page 231.
Parameters:
<StepSize> f
max
is specified in the data sheet.
Range: 1 to fMAX
*RST: 0.1 x span
Default unit: Hz
Example:
FREQ:CENT 100 MHz
FREQ:CENT:STEP 10 MHz
FREQ:CENT UP
Sets the center frequency to 110 MHz.
Manual operation: See "Center Frequency Stepsize" on page 93
[SENSe:]FREQuency:CENTer:STEP:LINK <CouplingType>
This command couples and decouples the center frequency step size to the span or
the resolution bandwidth.