Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
236User Manual 1173.9240.02 ─ 17
This function is not available if the Digital Baseband Interface (R&S FSW-B17) is
active.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
Usage: SCPI confirmed
Manual operation: See "Attenuation Mode / Value" on page 88
INPut:
EATT <Attenuation>
This command defines an electronic attenuation manually. Automatic mode must be
switched off (INP:EATT:AUTO OFF, see INPut:EATT:AUTO on page 236).
If the current reference level is not compatible with an attenuation that has been set
manually, the command also adjusts the reference level.
This command is only available with option R&S FSW-B25.
It is not available if R&S FSW-B17 is active.
Parameters:
<Attenuation> attenuation in dB
Range: see data sheet
Increment: 1 dB
*RST: 0 dB (OFF)
Example:
INP:EATT:AUTO OFF
INP:EATT 10 dB
Manual operation: See "Using Electronic Attenuation (Option B25)" on page 88
INPut:EATT:AUTO <State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
This command is only available with option R&S FSW-B25.
It is not available if R&S FSW-B17 is active.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
INP:EATT:AUTO OFF
Manual operation: See "Using Electronic Attenuation (Option B25)" on page 88
Configuring the Measurement