Configuration
R&S
®
FSW-K7
65User Manual 1173.9240.02 ─ 17
Edit Table
Opens the "Edit Conversion loss table" dialog box to edit the selected conversion loss
table. For details on table configuration see "Creating and Editing Conversion Loss
Tables" on page 65.
Remote command:
[SENSe:]CORRection:CVL:SELect on page 195
Delete Table
Deletes the currently selected conversion loss table after you confirm the action.
Remote command:
[SENSe:]CORRection:CVL:CLEAr on page 192
Import Table
Imports a stored conversion loss table from any directory and copies it to the instru-
ment's C:\r_s\instr\user\cvl\ directory. It can then be assigned for use for a
specific frequency range (see "Conversion loss" on page 61).
Creating and Editing Conversion Loss Tables
Conversion loss tables can be defined and edited in the "Edit conversion loss table"
dialog box which is displayed when you select the "New Table" button in the "External
Mixer > Conversion loss table" settings.
A preview pane displays the current configuration of the conversion loss function as
described by the position/value entries.
Input and Frontend Settings