Applications and Operating Modes
R&S
®
FSW
105User Manual 1173.9411.02 ─ 19
New Channel
The applications selected on this tab are started in a new channel, i.e. a new tab in the
display.
Remote command:
INSTrument:CREate[:NEW] on page 683
INSTrument[:SELect] on page 686
Replace Current Channel
The applications selected on this tab are started in the currently displayed channel,
replacing the current application.
Remote command:
INSTrument:CREate:REPLace on page 684
Duplicate Current Channel
The currently active channel can be duplicated, i.e. a new measurement channel of the
same type and with the identical measurement settings is started. The name of the
new channel is the same as the copied channel, extended by a consecutive number
(e.g. "Spectrum" -> "Spectrum 2").
This command is not available if the MSRA Master channel is selected.
Remote command:
INSTrument:CREate:DUPLicate on page 683
4.5 Running a Sequence of Measurements
Only one measurement can be performed at any time, namely the one in the currently
active channel. However, in order to perform the configured measurements consecu-
tively, a Sequencer function is provided.
The Sequencer function is slightly different in MSRA or MSRT operating mode. For
details see the R&S FSW MSRA User Manual or the R&S FSW Realtime Spectrum
Application and MSRT Operating Mode User Manual.
The Sequencer Concept....................................................................................... 105
Sequencer Settings...............................................................................................107
How to Set Up the Sequencer...............................................................................108
4.5.1 The Sequencer Concept
The instrument can only be in one specific channel at any time. Thus, only one mea-
surement can be performed at any time, namely the one in the currently active chan-
nel. However, in order to perform the configured measurements consecutively, a
Sequencer function is provided, which changes the application of the instrument as
required. If activated, the measurements configured in the currently active channels
are performed one after the other in the order of the tabs. For each individual measure-
Running a Sequence of Measurements