Remote Commands
R&S
®
FSW
729User Manual 1173.9411.02 ─ 19
CALC:MARK:FUNC:POW:SEL OBW
//Activates occupied bandwidth measurement.
POW:BWID 99PCT
//Sets the percentage of power to 99%.
POW:ACH:BAND 21kHz
//Sets the channel bandwidth to 21 kHz.
POW:ACH:PRES OBW
//Optimizes the instrument settings according to the channel bandwidth.
POW:ACH:PRES:RLEV
//Determines the ideal reference level for the measurement.
DET APE
//Sets the trace detector to positive peak.
//--------------Performing the Measurement-----
INIT:CONT OFF
//Selects single sweep mode.
INIT;*WAI
//Initiates a new measurement and waits until the sweep has finished.
//---------------Retrieving Results-------------
CALC:MARK:FUNC:POW:RES? OBW
//Returns the occupied bandwidth.
11.5.6 Measuring the Spectrum Emission Mask
All remote control commands specific to spectrum emission mask measurements are
described here.
See also chapter 11.5.2, "Configuring Power Measurements", on page 695.
Managing Measurement Configurations............................................................... 729
Controlling the Measurement................................................................................731
Configuring a Multi-SEM Measurement................................................................ 732
Configuring a Sweep List...................................................................................... 732
Configuring the Reference Range.........................................................................747
Configuring the Power Classes.............................................................................749
Configuring MSR SEM Measurements................................................................. 754
Configuring the List Evaluation............................................................................. 760
Performing an SEM Measurement........................................................................761
Retrieving Results.................................................................................................761
Example: SEM Measurement............................................................................... 762
11.5.6.1 Managing Measurement Configurations
The following commands control measurement configurations for SEM measurements.
Configuring and Performing Measurements