Remote Commands
R&S
®
FSW
900User Manual 1173.9411.02 ─ 19
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET2:MTIM:AVER ON
Activates manual averaging.
Manual operation: See "Meas Time/Average" on page 326
[SENSe:
]PMETer<p>:ROFFset[:STATe] <State>
This command includes or excludes the reference level offset of the analyzer for power
sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | 1
Includes the reference level offset in the results.
OFF | 0
Ignores the reference level offset.
*RST: 1
Example:
PMET2:ROFF OFF
Takes no offset into account for the measured power.
Manual operation: See "Use Ref Lev Offset" on page 327
[SENSe:]PMETer<p>[:STATe] <State>
This command turns a power sensor on and off.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET1 ON
Switches the power sensor measurements on.
Manual operation: See "State" on page 325
See "Select" on page 325