Remote Commands
R&S
®
FSW
918User Manual 1173.9411.02 ─ 19
DISPlay[:WINDow<n>]:SGRam|SPECtrogram:COLor:LOWer <Percentage>
This command defines the starting point of the color map.
Parameters:
<Percentage> Statistical frequency percentage.
Range: 0 to 66
*RST: 0
Default unit: %
Example:
DISP:WIND:SGR:COL:LOW 10
Sets the start of the color map to 10%.
Manual operation: See "Start / Stop" on page 476
DISPlay[:
WINDow<n>]:SGRam|SPECtrogram:COLor:SHAPe <Shape>
This command defines the shape and focus of the color curve for the spectrogram
result display.
Parameters:
<Shape> Shape of the color curve.
Range: -1 to 1
*RST: 0
Manual operation: See "Shape" on page 476
DISPlay[:WINDow<n>]:SGRam|SPECtrogram:COLor:UPPer <Percentage>
This command defines the end point of the color map.
Parameters:
<Percentage> Statistical frequency percentage.
Range: 0 to 66
*RST: 0
Default unit: %
Example:
DISP:WIND:SGR:COL:UPP 95
Sets the start of the color map to 95%.
Manual operation: See "Start / Stop" on page 476
DISPlay[:WINDow<n>]:SGRam|SPECtrogram:COLor[:STYLe] <ColorScheme>
This command selects the color scheme.