The R&S FSW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Parameters:
<Threshold> Numeric value.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 982.
*RST: -200 dBm
Manual operation: See "Threshold" on page 532
CALCulate:
LIMit<k>:NAME <Name>
This command selects a limit line that already exists or defines a name for a new limit
line.
Parameters:
<Name> String containing the limit line name.
*RST: REM1 to REM8 for lines 1 to 8
Manual operation: See "Name" on page 532
CALCulate:LIMit<k>:UNIT <Unit>
This command defines the unit of a limit line.
Parameters:
<Unit> DBM | DBPW | WATT | DBUV | DBMV | VOLT | DBUA |
AMPere | DB | DBUV_M | DBUA_M | (unitless)
If you select dB as the limit line unit, the command automatically
turns the limit line into a relative limit line.
*RST: DBM
Manual operation: See "Y-Axis" on page 533
CALCulate:LIMit<k>:UPPer[:DATA] <LimitLinePoints>
This command defines the vertical definition points of an upper limit line.
Parameters:
<LimitLinePoints> Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with CALCulate:
LIMit<k>:CONTrol[:DATA]. If not, the R&S FSW either adds
missing values or ignores surplus values.
The unit depends on CALCulate:LIMit<k>:UNIT
on page 982.
*RST: Limit line state is OFF
Usage: SCPI confirmed