Remote Commands
R&S
®
FSW
830User Manual 1173.9411.02 ─ 19
11.7 Setting Basic Measurement Parameters
All commands that set measurement-independent parameters are described here.
● Defining the Frequency and Span.........................................................................830
● Configuring Bandwidth and Sweep Settings.........................................................836
● Configuring the Vertical Axis (Amplitude, Scaling)................................................844
● Configuring Triggered and Gated Measurements.................................................850
● Adjusting Settings Automatically...........................................................................860
● Configuring the Data Input and Output................................................................. 864
11.7.1 Defining the Frequency and Span
The commands required to configure the frequency and span settings in a remote envi-
ronment are described here. The tasks for manual operation are described in chap-
ter 6.3, "Frequency and Span Configuration", on page 384.
● Defining the Frequency Range............................................................................. 830
● Configuring Signal Tracking..................................................................................834
11.7.1.1 Defining the Frequency Range
CALCulate<n>:MARKer<m>:FUNCtion:CENTer................................................................830
CALCulate<n>:MARKer<m>:FUNCtion:CSTep................................................................. 831
DISPlay[:WINDow<n>]:TRACe:X:SPACing.......................................................................831
[SENSe:]FREQuency:CENTer.........................................................................................831
[SENSe:]FREQuency:CENTer:STEP............................................................................... 832
[SENSe:]FREQuency:CENTer:STEP:AUTO..................................................................... 832
[SENSe:]FREQuency:CENTer:STEP:LINK....................................................................... 832
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor...........................................................833
[SENSe:]FREQuency:OFFSet.........................................................................................833
[SENSe:]FREQuency:SPAN............................................................................................833
[SENSe:]FREQuency:SPAN:FULL...................................................................................834
[SENSe:]FREQuency:STARt...........................................................................................834
[SENSe:]FREQuency:STOP............................................................................................834
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
This command matches the center frequency to the frequency of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Example:
CALC:MARK2:FUNC:CENT
Sets the center frequency to the frequency of marker 2.
Usage: Event
Manual operation: See "Harmonic Distortion" on page 114
See "Center Frequency = Marker Frequency" on page 507