Remote Commands for the I/Q Analyzer
R&S
®
FSWP
142User Manual 1177.5856.02 ─ 07
<WindowIndex> numeric value
Index of the window.
Example:
LAY:CAT?
Result:
'2',2,'1',1
Two windows are displayed, named '2' (at the top or left), and '1'
(at the bottom or right).
Usage: Query only
LAYout:IDENtify[:WINDow]? <WindowName>
This command queries the index of a particular display window in the active channel.
Note: to query the name of a particular window, use the LAYout:WINDow<n>:
IDENtify? query.
Query parameters:
<WindowName> String containing the name of a window.
Return values:
<WindowIndex> Index number of the window.
Example:
LAY:WIND:IDEN? '2'
Queries the index of the result display named '2'.
Response:
2
Usage: Query only
LAYout:REMove[:WINDow] <WindowName>
Setting parameters:
<WindowName>
Usage: Setting only
LAYout:REPLace[:WINDow] <WindowName>,<WindowType>
This command replaces the window type (for example from "Diagram" to "Result Sum-
mary") of an already existing window in the active channel while keeping its position,
index and window name.
To add a new window, use the LAYout:ADD[:WINDow]? command.
Setting parameters:
<WindowName> String containing the name of the existing window.
By default, the name of a window is the same as its index. To
determine the name and index of all active windows in the active
channel, use the LAYout:CATalog[:WINDow]? query.
<WindowType> Type of result display you want to use in the existing window.
See LAYout:ADD[:WINDow]? on page 140 for a list of availa-
ble window types.
Configuring I/Q Analyzer Measurements