Remote Commands for the I/Q Analyzer
R&S
®
FSWP
94User Manual 1177.5856.02 ─ 07
[SENSe:]SWEep[:WINDow<n>]:POINts <SweepPoints>
This command defines the number of measurement points to analyze after a measure-
ment.
Suffix:
<n>
.
Example:
SWE:POIN 251
Manual operation: See "Sweep Points" on page 63
[SENSe<n>:]SWEep:TIME <Time>
This command defines the measurement time. It automatically decouples the time from
any other settings.
Suffix:
<n>
.
irrelevant
Parameters:
<Time> refer to data sheet
*RST: depends on current settings (determined automati-
cally)
Default unit: S
Manual operation: See " Meas Time " on page 60
SYSTem:SEQuencer <State>
This command turns the Sequencer on and off. The Sequencer must be active before
any other Sequencer commands (INIT:SEQ...) are executed, otherwise an error will
occur.
A detailed programming example is provided in the "Operating Modes" chapter in the
R&S FSWP User Manual.
Parameters:
<State> ON | OFF | 0 | 1
ON | 1
The Sequencer is activated and a sequential measurement is
started immediately.
OFF | 0
The Sequencer is deactivated. Any running sequential measure-
ments are stopped. Further Sequencer commands
(INIT:SEQ...) are not available.
*RST: 0