Acquisition and waveform setupMXO 4 Series
123User Manual 1335.5337.02 ─ 07
Bandwidth
Sets the bandwidth limit. The specified bandwidth indicates the range of frequencies
that the instrument can acquire and display accurately with less than 3 dB attenuation.
Frequencies above the limit are removed from the signal, and noise is reduced.
The selected bandwidth is shown on the signal icon.
For basic information, see also: "Bandwidth" on page 119.
"Full"
At full bandwidth, all frequencies in the instrument's frequency range
are acquired and displayed. Full bandwidth is used for most applica-
tions. However, at full bandwidth, the displayed bandwidth can be
less than the instrument bandwidth depending on the number of
active channels and other settings.
"xx MHz"
Frequencies above the selected limit are removed to reduce noise.
Remote command:
CHANnel<ch>:BANDwidth on page 512
Probe type, Probe bandwidth
Shows the type of the connected probe and its bandwidth. The probe is recognized
automatically, or selected in the "Probe" tab. See Chapter 6.4.1, "Common probe set-
tings", on page 125.
Effective bandwidth
Shows the effective bandwidth of probe and oscilloscope system.
6.3.4 Probe settings
See Chapter 6.4, "Probes", on page 125.
6.3.5 Other vertical settings
Access: "Menu" > "Vertical" > "Other" tab
Vertical setup