TriggerMXO 4 Series
159User Manual 1335.5337.02 ─ 07
"Shorter"
Triggers on intervals shorter than the given "Interv. width".
"Longer"
Triggers on intervals longer than the given "Interv. width".
Remote command:
TRIGger:EVENt<ev>:INTerval:RANGe on page 556
Interv. width
Sets the time between two pulses for the interval trigger.
Remote command:
TRIGger:EVENt<ev>:INTerval:WIDTh on page 557
±Delta
Defines a range around the "Interv. width" value.
Remote command:
TRIGger:EVENt<ev>:INTerval:DELTa on page 555
7.4.9 Slew rate trigger
Access: "Menu" > "Trigger" > "Setup" tab > "Type = Slew rate"
The slew rate trigger is also known as transition trigger. It triggers if the transition time
from the lower to higher voltage level (or vice versa) is shorter or longer as defined, or
outside or inside a specified time range.
The slew rate trigger finds slew rates faster than expected or permissible to avoid over-
shooting and other interfering effects. It also detects slow edges violating the timing in
pulse series.
This section describes the trigger-type specific settings. For settings that apply to all
trigger types, see Chapter 7.2, "Common trigger settings", on page 142.
Trigger types