Note that reference waveforms can be loaded only from REF files. CSV files are
meant for further processing in other applications.
5. To save the waveform, tap "Save".
6. Enter a filename and select the directory. The file type is defined according to the
selection in the previous steps.
The source settings of the reference waveform and the current scaling settings are
stored to the specified file.
To load a reference waveform
Note: Reference waveforms can be loaded only from REF files.
1. Press the [Ref] key.
2. Select the tab for the reference waveform that you want to load ("R1/2/3/4").
3. In the "Setup" tab, select "Recall".
4. Select the file from the file selection dialog box.
5. To load the waveform from the specified file, tap "Open".
The selected waveform is loaded as the specified reference waveform.
8.4.2 Settings for reference waveforms
To compare waveforms and analyze differences between waveforms, you can use up
to four reference waveforms R1 to R4.
Each reference waveform has its own memory on the instrument. You can also save
an unlimited number of reference waveforms and load them for further use.
The display of a reference waveform is independent from the display of the source
waveform; you can move, stretch and compress the curve vertically and horizontally.
8.4.2.1 Reference waveform setup
Access: [Ref] key > "Setup" tab