Waveform analysisMXO 4 Series
206User Manual 1335.5337.02 ─ 07
Like vertical offset of a channel waveform, the offset of a reference waveform is sub-
tracted from the measured value. Negative values shift the waveform up, positive val-
ues shift it down.
Remote command:
REFCurve<rc>:RESCale:VERTical:OFFSet on page 596
Horizontal
Stretching and offset change the display of the waveform independent of the horizontal
settings of the source waveform and of the horizontal diagram settings.
Mode ← Horizontal
Selects the type of horizontal settings:
"Original scal-
ing"
Horizontal scaling and reference point of the source waveform are
used.
"Adjust to X
Axis"
The current horizontal settings of the diagram are used.
Remote command:
REFCurve<rc>:HMODe on page 594
Enable ← Horizontal
If enabled, the horizontal offset and factor are applied to the reference waveform.
Remote command:
REFCurve<rc>:RESCale:HORizontal:STATe on page 595
Scale factor ← Horizontal
Sets the horizontal scale factor. A factor greater than 1 stretches the waveform hori-
zontally, a factor lower than 1 compresses the curve.
Remote command:
REFCurve<rc>:RESCale:HORizontal:FACTor on page 594
Horizontal offset ← Horizontal
Moves the waveform horizontally. Enter a value with a time unit suitable for the time
scale of the diagram.
Positive values shift the waveform to the right, negative values shift it to the left.
Remote command:
REFCurve<rc>:RESCale:HORizontal:OFFSet on page 595
Reference waveforms