ApplicationsMXO 4 Series
261User Manual 1335.5337.02 ─ 07
Remote command:
FRANalysis:FREQuency:STARt on page 662
FRANalysis:FREQuency:STOP on page 662
Points
Selects, if the number of points are measured as total or per decade and sets the num-
ber of points.
Remote command:
FRANalysis:POINts:TOTal on page 665
FRANalysis:POINts:MODE on page 665
FRANalysis:POINts:LOGarithmic on page 664
Amplitude
Sets a fixed amplitude for the frequency response analysis.
Remote command:
FRANalysis:GENerator:AMPLitude on page 662
User load
Selects the generator voltage display for 50Ω or high impedance load.
Remote command:
FRANalysis:GENerator:LOAD on page 662
Amplitude mode
Selects, if the amplitude is a constant value ("Amplitude") or is defined as an amplitude
profile.
Remote command:
FRANalysis:POINts:MODE on page 665
Amplitude profile
Opens a dialog to set the amplitude profile. See Chapter 11.1.3.2, "Amplitude profile",
on page 262.
You can then define different amplitudes for different frequencies. The amplitude profile
is useful when testing sensitive circuits, where the amplitude gets too high. In this case
distortion might occur.
If this function is enabled, a green colored diagram of the amplitudes for the different
frequencies can be displayed on the screen.
Remote command:
FRANalysis:AMPLitude:MODE on page 661
FRANalysis:AMPLitude:ENABle on page 670
Run
Starts the frequency response analysis.
Remote command:
FRANalysis:STATe on page 661
Repeat
Repeats the measurement, using the same parameters.
Frequency response analysis (option R&S
MXO4-K36)