ApplicationsMXO 4 Series
276User Manual 1335.5337.02 ─ 07
a) Connect the positive (+) signal socket of the differential voltage probe to the
line of the AC input.
b) Connect the negative (-) signal socket of the differential voltage probe to the
neutral of the AC input.
c) Connect the current probe to the line of the AC input.
5. Select the correct channels for the "Voltage" and the "Current" probes.
6. Set the "Fund. Freq" according to your signal.
7. If necessary, enable "Statistics".
On the screen, you can see the measurement waveforms of the current, the volt-
age and the power. Also, the result box with numeric measurement results is
shown. For details, see "Power quality results" on page 278.
Power quality settings
Voltage
Selects the channel for the voltage source.
Power analysis (option R&S
MXO4-K31)