Getting StartedMXO 4 Series
33User Manual 1335.5337.02 ─ 07
If the monitor provides touch functionality, an additional connection can be required,
e.g. a USB connection. Refer to the documentation of your monitor.
The oscilloscope outputs a Full HD image with 1920x1080 pixel. The instrument's dis-
play is duplicated. If the screen resolution of the monitor is set higher than the instru-
ment's resolution, the application window uses a 1920x1080 area of the monitor dis-
play. For full screen display, adjust the monitor's screen resolution.
3.1.9 Switching on or off
The instrument is switched on or off with the power switch and the [Power] key. The
[Power] key is located in the bottom left corner of the front panel. The power switch is
located at the rear panel of the instrument.
Table 3-1: Overview of power states
Status LED Power switch
Off
(unlighted)
[0] (off)
Standby
yellow
[I] (on)
Ready
green
[I] (on)
To switch on the product
The product is off but connected to power.
1. Set the switch on the power supply to position [I].
The power key on the front panel lights up.
2. Press the [Power] key on the front panel.
The instrument performs a system check, boots the operating system, and then
starts the MXO 4 firmware.
The [Power] key turns green and the illuminated keys on the front panel light up. If
the previous session was terminated regularly, the oscilloscope uses the last set-
tings.
Before you start measurements, be sure to comply with the warm-up phase specified
in the specifications document.
To shut down the product
The product is in the ready state.
► Press the [Power] key.
All current settings are saved, and the software shuts down. The [Power] key turns
yellow. The standby power supplies only the power switch circuits.
The
"Power" icon in the "Menu" shuts down only the firmware application. To shut
down the instrument completely, use the [Power] key.
Remote command for firmware shutdown: SYSTem:EXIT on page 486.
Preparing for use