EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - I²C Trigger Settings

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysisMXO 4 Series
344User Manual 1335.5337.02 ─ 07
"Data"
The data setup consists of a comparison condition and one or two
data patterns.
Remote command:
SBUS<sb>:I2C:FILTer:FRAMe<fr>:FLD<fl>:DMAX
on page 735
SBUS<sb>:I2C:FILTer:DMAX on page 735
SBUS<sb>:I2C:FILTer:FRAMe<fr>:FLD<fl>:DMIN
on page 735
SBUS<sb>:I2C:FILTer:DMIN on page 735
SBUS<sb>:I2C:FILTer:FRAMe<fr>:FLD<fl>:DOPerator
on page 736
SBUS<sb>:I2C:FILTer:DOPerator on page 736
"Index"
The index setup consists of a comparison condition and one or two
index values.
Remote command:
SBUS<sb>:I2C:FILTer:FRAMe<fr>:FLD<fl>:IMAX
on page 736
SBUS<sb>:I2C:FILTer:IMAX on page 736
SBUS<sb>:I2C:FILTer:FRAMe<fr>:FLD<fl>:IMIN
on page 737
SBUS<sb>:I2C:FILTer:IMIN on page 737
SBUS<sb>:I2C:FILTer:FRAMe<fr>:FLD<fl>:IOPerator
on page 737
SBUS<sb>:I2C:FILTer:IOPerator on page 737
Error type
Enables filtering on the selected error type.
Remote command:
SBUS<sb>:I2C:FILTer:ERENable on page 736
SBUS<sb>:I2C:FILTer:ERRor<n>:ENABle on page 736
13.3.4 I²C trigger settings
Access: [Apps] > "Protocol" > "I²C" > "Shortcuts" > "Setup trigger"
I²C (option R&S MXO4-K510)

Table of Contents