EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 382

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysisMXO 4 Series
382User Manual 1335.5337.02 ─ 07
"XLFF"
XL frame format: supports a length of 29 bit for the CAN identifier.
"Error"
When a node recognizes an error, it cancels transmission by sending
an error frame.
The ID type is irrelevant for error frames.
"Overload"
When a node needs a delay between data and/or remote frames, it
sends and overload frame.
The ID type is irrelevant for overload frames.
Remote command:
TRIGger:CAN:FTYPe on page 770
Identifier
Specifies the identifier pattern.
"Condition"
Sets the comparison condition to a specific value or a range.
"Value"/"From"
Specifies the value or sets the start value of a range.
"To"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger:CAN:ICONdition on page 771
TRIGger:CAN:IMAX on page 771
TRIGger:CAN:IMIN on page 771
DLC
The data length code (DLC) defines the number of data bytes to be found. You can set
a fixed value for the DLC or define a minimum value.
CAN (option R&S
MXO4-K520)

Table of Contents

Related product manuals