Protocol analysisMXO 4 Series
395User Manual 1335.5337.02 ─ 07
Remote command:
SBUS<sb>:LIN:DATA:SOURce on page 796
Polarity
Defines the idle state of the bus. The idle state is the recessive state and corresponds
to a logic high.
Remote command:
SBUS<sb>:LIN:POLarity on page 797
Bit rate
Sets the number of transmitted bits per second. The maximum bit rate for LIN is 20
kbit/s.
If the LIN standard is "J2602", the bit rate is 10.417 kbit/s and cannot be changed.
Remote command:
SBUS<sb>:LIN:BITRate on page 796
13.6.1.2 Threshold settings
Access: [Apps] key > "Protocol" tab > "LIN" > "Thresholds".
Threshold
Sets the threshold for the data channel. Enter the value directly in the fields.
Additional to the threshold, you can also set a hysteresis.
LIN (option R&S
MXO4-K520)