Getting StartedMXO 4 Series
40User Manual 1335.5337.02 ─ 07
[Level]
The rotary knob sets the trigger level for all trigger types that require one level. For trig-
ger types with two levels, the knob sets the lower level. Turn clockwise to move up the
trigger level. Press the knob to set the trigger level to 50% of the signal amplitude.
Remote command:
TRIGger:EVENt<ev>:LEVel<n>[:VALue] on page 541
[Run / Stop]
Starts and stops the continuous acquisition. The [Run / Stop] key lights green when the
acquisition is running. When the acquisition is stopped, the key lights red.
Remote command:
RUN on page 505
STOP on page 505
[Single]
Starts a defined number of acquisitions. The [Single] key lights green when the acquisi-
tion is running. When the acquisition is stopped, the key lights red.
Press the key again to stop a running acquisition. To set the number of acquisitions,
select "Menu" > "Acquisition", and set "N-single/Avg count".
Remote command:
SINGle on page 505
[Auto Norm]
Toggles the trigger mode between "Auto" and "Normal". The current setting is shown
on the trigger label.
Auto
The instrument triggers repeatedly after a time interval if the trigger
conditions are not fulfilled. If a real trigger occurs, it takes prece-
dence. The time interval depends on the timebase.
Norm
The instrument acquires a waveform only if a trigger occurs.
The "Norm" indicator above the key turns green.
Remote command:
TRIGger:MODE on page 565
[Source]
Toggles the trigger source: C1, C2, C3, etc. The key lights up in the color of the
selected channel.