Waveform generator (option R&S MXO4-B6)MXO 4 Series
439User Manual 1335.5337.02 ─ 07
Modulation state
Enables or disables modulation.
Modulation is available only for "Function type" = "Sine"/"Square".
Remote command:
WGENerator<wg>:MODulation[:STATe] on page 863
Modulation type
Selects the modulation type, which defines how the carrier signal is modified.
"AM"
Amplitude modulation. The amplitude of the carrier signal is varied
according to the modulation signal.
"FM"
Frequency modulation. The frequency of the carrier signal is varied
according to the modulation signal.
"PWM"
Pulse width modulation. The time for which the signal is in a high
state is varied according to the modulation signal.
"FSK"
Frequency shift keying (FSK) modulation. The signal frequency
switches between "Frequency 1" and "Frequency 2" at a "FSK rate".
Remote command:
WGENerator<wg>:MODulation:TYPE on page 870
Signal type
Selects the type of the modulating signal for AM, FM or PWM modulation.
Remote command:
WGENerator<wg>:MODulation:AM[:FUNCtion] on page 864
WGENerator<wg>:MODulation:FM[:FUNCtion] on page 866
WGENerator<wg>:MODulation:PWM[:FUNCtion] on page 870