3.2.5 Checking the functionality
To check if the instrument works correctly, you can use the probe compensation signal
and check the displayed signal.
1. Perform a self-alignment of the instrument:
a) Warm up the instrument. The minimum warm-up time is indicated in the specifi-
cations document.
b) Remove all probes from the input connectors.
c) Open the "Menu", and select "Settings" > "Maintenance" > "Alignment".
d) Tap "Start Alignment". Wait until the alignment has finished successfully.
2.
Connect the probe's ground connector to the ground pin , and the tip to the
square wave pin .
3. Press the [Preset] key.
4. Connect the probe to the input connector [C1].
The instrument recognizes the probe, and a signal is displayed in the diagram.
5. Tap the signal icon of C1 to open the vertical settings.
6. In the "Vertical" > "Setup" dialog, adjust the following:
"Vertical scale" = 600 mV/div
"Offset" = 900 mV
7. Turn the horizontal [Scale] knob to set the horizontal scale to 1 ms/div.
8. Press the [Source] key to set the trigger source to C1. The color of the key indi-
cates the selected channel.