Remote control commandsMXO 4 Series
499User Manual 1335.5337.02 ─ 07
DISPlay:PERSistence[:STATe] <State>
If enabled, each new data point in the diagram area remains on the screen for the
duration defined using DISPlay:PERSistence:TIME, or as long as DISPlay:
PERSistence:INFinite is enabled.
If disabled, the signal value is only displayed as long as it actually occurs.
Parameters:
<State> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Enable" on page 90
17.7.5.2
Signal
DISPlay:INTensity.......................................................................................................... 499
DISPlay:DIAGram:STYLe............................................................................................... 499
DISPlay:INTensity <Intensity>
The intensity determines the strength of the waveform line in the diagram. Enter a per-
centage between 0 (not visible) and 100% (strong). The default value is 50%.
Parameters:
<Intensity> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Intensity" on page 91
DISPlay:DIAGram:STYLe <Style>
Selects the style in which the waveform is displayed.
Parameters:
<Style> VECTors | DOTS
VECTors
The individual data points are connected by a line.
DOTS
Only the individual data points are displayed.
*RST: VECTors
Usage: Asynchronous command
Manual operation: See "Waveform style" on page 91