Remote control commandsMXO 4 Series
500User Manual 1335.5337.02 ─ 07
17.7.5.3 Backlight
DISPlay:BACKlight[:BRIGtness].......................................................................................500
DISPlay:BACKlight:DIMMing........................................................................................... 500
DISPlay:BACKlight[:BRIGtness] <LCDIntensity>
Sets the background luminosity of the touchscreen.
Parameters:
<LCDIntensity> Range: 5 to 100
Increment: 1
*RST: 90
Default unit: %
Usage: Asynchronous command
Manual operation: See "Brightness" on page 92
DISPlay:BACKlight:DIMMing
<UsrActivityTout>
Selects a time, after which the monitor brightness is reduced, if the instrument was
inactive. Remote control of the instrument is also considered as an activity.
Parameters:
<UsrActivityTout> OFF | T1Minute | T2Minutes | T3Minutes | T5Minutes |
T10Minutes | T15Minutes | T20Minutes | T25Minutes |
T30Minutes | T45Minutes | T1Hour | T2Hours | T3Hours |
T4Hours
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Dimm on user inactivity" on page 92
17.7.5.4 Clear results
DISPlay:CLR
Deletes all measurement results including all waveforms and statistics.
Usage:
Setting only
Asynchronous command
Manual operation: See "Clear" on page 62