Remote control commandsMXO 4 Series
504User Manual 1335.5337.02 ─ 07
Manual operation: See "Date, Time, Overall alignment state" on page 98
CALibration:RESult?
Returns the result of the last self-alignment and the current alignment status. In remote
mode, *CAL? provides more detailed information.
Return values:
<ResultState> PASSed | FAILed | NOALigndata
*RST: FAILed
Usage: Query only
Asynchronous command
Manual operation: See "Date, Time, Overall alignment state" on page 98
SERVice:REPort
Creates a service report.
The service report is a ZIP file with a complete bug report, all relevant setup informa-
tion, reporting and log files, alignment files, and the instrument configuration.
If a USB flash drive is connected, the report is saved on the USB flash drive. Other-
wise, the report is saved in the user data folder /home/storage/userData.
Usage:
Event
Asynchronous command
Manual operation: See "Create report" on page 100
SYSTem:APUP <AutoPowerUp>
If enabled, the instrument powers up automatically when it is connected to the mains
voltage.
Parameters:
<AutoPowerUp> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Auto power up" on page 100
Instrument setup