Remote control commandsMXO 4 Series
507User Manual 1335.5337.02 ─ 07
TIMebase:HORizontal:POSition <Position>
Defines the time distance between the reference point and the trigger point, which is
the zero point of the diagram. The horizontal position is also known as trigger offset.
Parameters:
<Position> Range: Device specific to 1E+26
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "[Position]" on page 41
TIMebase:REFerence <RescaleCtrPos>
Sets the position of the reference point in % of the screen. It defines which part of the
waveform is shown.
Parameters:
<RescaleCtrPos> Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Reference point" on page 107
17.8.3 Roll mode
TIMebase:ROLL:ENABle................................................................................................ 507
TIMebase:ROLL:MTIMe..................................................................................................508
ACQuire:ROLLmode:OSCapture..................................................................................... 508
ACQuire:ROLLmode:POINts?......................................................................................... 508
TIMebase:ROLL:STATe?.................................................................................................508
TIMebase:ROLL:ENABle <Mode>
Selects, if the roll mode is set automatically by the instrument or if it is turned off.
Parameters:
<Mode> AUTO | OFF
*RST: AUTO
Usage: Asynchronous command
Manual operation: See "Mode" on page 108
Acquisition and setup