Remote control commandsMXO 4 Series
510User Manual 1335.5337.02 ─ 07
Parameters:
<Scale> Range: 0.001 to 1
Increment: 0.001
*RST: 0.05
Default unit: Depends on the connected probe
Usage: Asynchronous command
Manual operation: See "[Scale]" on page 43
CHANnel<ch>:RANGe <Range>
Sets the voltage range across the 10 vertical divisions of the diagram. The command is
an alternative to CHANnel<ch>:SCALe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Range> Range: 0.01 to 10
Increment: 0.01
*RST: 0.5
Default unit: V/div
Usage: Asynchronous command
CHANnel<ch>:OFFSet <Offset>
Sets the offset voltage, which corrects an offset-affected signal. The vertical center of
the selected channel is shifted by the offset value and the signal is repositioned within
the diagram.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Offset> Range: -1 to 1
Increment: 0.01
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "[Position]" on page 43
CHANnel<ch>:POSition <Position>
Moves the selected signal up or down in the diagram. While the offset sets a voltage,
position is a graphical setting given in divisions. The visual effect is the same as for off-