Remote control commandsMXO 4 Series
524User Manual 1335.5337.02 ─ 07
PROBe<ch>:SETup:OFFSet:AZERo
Measures the zero error. Short the signal pin and the ground pin together and connect
them to the ground of the DUT before sending the command.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Usage: Event
Asynchronous command
Manual operation: See "AutoZero, Use AutoZero" on page 128
PROBe<ch>:SETup:OFFSet:USEautozero
<UseAutoZeroOffset>
Includes the AutoZero offset in measurement results. The AutoZero error is detected
with PROBe<ch>:SETup:OFFSet:AZERo.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<UseAutoZeroOffset> ON | OFF
*RST: OFF
Example:
PROB2:SET:OFFS:AZER
PROB2:SET:OFFS:USE ON
Detects the zero error and uses it for correction of measurement
results.
Usage: Asynchronous command
Manual operation: See "AutoZero, Use AutoZero" on page 128
PROBe<ch>:SETup:NAME?
Queries the name of the probe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Return values:
<Name> Name string
Usage: Query only
Asynchronous command
Manual operation: See "Predefined probe, name and type of the probe"
on page 126
PROBe<ch>:SETup:STATe?
Queries if the probe at the specified input channel is active (detected) or not active (not
detected). To switch the probe on, use CHANnel<ch>:STATe.