Remote control commandsMXO 4 Series
526User Manual 1335.5337.02 ─ 07
PROBe<ch>:SETup:MODE <Mode>
The micro button is located on the probe head. Pressing this button, you initiate an
action on the instrument directly from the probe. The button is disabled during internal
automatic processes, for example, during self-alignment, autoset, and level detection.
Select the action that you want to start from the probe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Mode> RCONtinuous | RSINgle | AUToset | AZERo | OTMean | SITFile |
NOACtion | FTRiglevel | PRSetup
RCONtinuous
Run continuous: the acquisition is running as long as the probe
button is pressed.
RSINgle
Run single: starts a defined number of acquisitions (same as
[Single] key).
AUToset
Starts the autoset procedure.
AZERo
AutoZero: performs an automatic correction of the zero error.
OTMean
Set offset to mean: performs an automatic compensation for a
DC component of the input signal.
SITFile
Save image to file: saves the display image in a file.
NOACtion
Nothing is started on pressing the micro button.
FTRiglevel
Sets the trigger level automatically to 0.5 * (MaxPeak – Min-
Peak). The function is not available for an external trigger
source.
PRSetup
Opens the "Probes Setup" dialog box.
*RST:
RCONtinuous
Usage: Asynchronous command
Manual operation: See "MicroButton" on page 130
PROBe<ch>:SETup:ACCoupling <ProbeCouplingAC>
Enables AC coupling in R&S RTZPR power rail probes, which removes DC and very
low-frequency components. The R&S RTZPR probe requires 50 Ω input termination,
for which the channel AC coupling is not available. The probe setting allows AC cou-
pling also at 50 Ω inputs.