Remote control commandsMXO 4 Series
532User Manual 1335.5337.02 ─ 07
PROBe<ch>:SETup:PRMode <MeasMode>
Sets the measurement mode of modular probes.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<MeasMode> DMODe | CMODe | PMODe | NMODe
DMODe
Differential mode input voltage (V
dm
), the voltage between the
positive and negative input terminal.
CMODe
Common mode input voltage (V
cm
), the mean voltage between
the positive and negative input terminal vs. ground.
PMODe
Positive single-ended input voltage (V
p
). the voltage between the
positive input terminal and ground.
NMODe
Negative single-ended input voltage (V
N
). the voltage between
the negative input terminal and ground.
*RST:
DMODe
Usage: Asynchronous command
PROBe<ch>:SETup:CMOFfset <CMOffset>
Sets the common-mode offset to compensate for a common DC voltage that is applied
to both input sockets (referenced to the ground socket). The setting is available for
Rohde & Schwarz differential probes and for modular probes in CM measurement
mode.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<CMOffset> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "CM offset" on page 132
PROBe<ch>:SETup:DMOFfset <DMOffset>
Sets the differential offset to compensate a DC voltage applied between the positive
(V
p
) and the negative (V
n
) input terminal at the probe tip.
Acquisition and setup