Remote control commandsMXO 4 Series
543User Manual 1335.5337.02 ─ 07
TRIGger:MEVents:SEQuence<se>:COUNt <Events>
Sets the number of B-trigger conditions to be fulfilled after an A-trigger. The last B-trig-
ger causes the trigger event. The waiting time for B-triggers can be restricted with a
reset condition: timeout or reset event.
Suffix:
<se>
.
1..3
2
Parameters:
<Events> Range: 1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "B event count" on page 145
TRIGger:MEVents:SEQuence<se>:DELay
<Delay>
Sets the time that the instrument waits after an A-trigger until it recognizes B-triggers.
Suffix:
<se>
.
1
Parameters:
<Delay> Range: 0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
Manual operation: See "Delay" on page 145
TRIGger:MEVents:SEQuence<se>:RESet:EVENt <State>
If enabled, the trigger sequence is restarted by the R-trigger condition if the specified
number of B-triggers does not occur before the R-trigger conditions are fulfilled.
Suffix:
<se>
.
3
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Enable reset event" on page 146
Trigger