Remote control commandsMXO 4 Series
545User Manual 1335.5337.02 ─ 07
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Slope" on page 147
TRIGger:ANEDge:LEVel <ExtTrigLev>
Sets the trigger level for the external trigger source.
Parameters:
<ExtTrigLev> Range: - 5 to 5
Default unit: V
Usage: Asynchronous command
Manual operation: See "Level" on page 143
TRIGger:ANEDge:COUPling
<Coupling>
Sets the connection of the external trigger signal, i.e. the input impedance and a termi-
nation. The coupling determines what part of the signal is used for triggering.
Parameters:
<Coupling> DC | DCLimit | AC
DC
Connection with 50 Ω termination, passes both DC and AC com-
ponents of the signal.
DCLimit
Connection with 1 MΩ termination, passes both DC and AC
components of the signal.
AC
Connection with 1 MΩ termination through DC capacitor,
removes DC and very low-frequency components. The wave-
form is centered on zero volts.
*RST:
DCLimit
Usage: Asynchronous command
Manual operation: See "Coupling" on page 148
TRIGger:ANEDge:FILTer <Filter>
Selects the filter mode for the external trigger signal.
Parameters:
<Filter> OFF | LFReject | RFReject
*RST: OFF
Usage: Asynchronous command