Remote control commandsMXO 4 Series
550User Manual 1335.5337.02 ─ 07
TRIGger:EVENt<ev>:LEVel<n>:RUNT:UPPer <Level>
Sets the upper voltage limit.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<Level> Range: -10 to 10
Increment: 0.001
*RST: 0.1
Default unit: V
Usage: Asynchronous command
Manual operation: See "Upper level" on page 153
TRIGger:EVENt<ev>:RUNT:DELTa
<WidthDelta>
Defines a range around the runt width specified using TRIGger:EVENt<ev>:RUNT:
WIDTh.
Available if TRIGger:EVENt<ev>:RUNT:RANGe is set to WITHin or OUTSide.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<WidthDelta> Range: 1E-10 to 864
Increment: 1E-07
*RST: 1E-10
Default unit: s
Usage: Asynchronous command
Manual operation: See "±Delta" on page 154
TRIGger:EVENt<ev>:RUNT:POLarity <Polarity>
Sets the polarity of a pulse, which is the direction of the first pulse slope.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Parameters:
<Polarity> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Polarity" on page 150
Trigger