EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Window Trigger

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
552User Manual 1335.5337.02 ─ 07
17.9.7 Window trigger
TRIGger:EVENt<ev>:LEVel<n>:WINDow:LOWer.............................................................. 552
TRIGger:EVENt<ev>:LEVel<n>:WINDow:UPPer............................................................... 552
TRIGger:EVENt<ev>:WINDow:DELTa.............................................................................. 552
TRIGger:EVENt<ev>:WINDow:RANGe............................................................................ 553
TRIGger:EVENt<ev>:WINDow:TIME................................................................................553
TRIGger:EVENt<ev>:WINDow:WIDTh............................................................................. 554
TRIGger:EVENt<ev>:LEVel<n>:WINDow:LOWer <Level>
Sets the lower voltage limit.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<Level> Range: -10 to 10
Increment: 0.001
*RST: -0.1
Default unit: V
Usage: Asynchronous command
Manual operation: See "Lower level" on page 156
TRIGger:EVENt<ev>:LEVel<n>:WINDow:UPPer <Level>
Sets the upper voltage limit.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<Level> Range: 0 to 1000
Increment: 1E-06
*RST: 0.1
Default unit: V
Usage: Asynchronous command
Manual operation: See "Upper level" on page 155
TRIGger:EVENt<ev>:WINDow:DELTa <WidthDelta>
Defines a range around the width value.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Trigger

Table of Contents