Remote control commandsMXO 4 Series
552User Manual 1335.5337.02 ─ 07
17.9.7 Window trigger
TRIGger:EVENt<ev>:LEVel<n>:WINDow:LOWer.............................................................. 552
TRIGger:EVENt<ev>:LEVel<n>:WINDow:UPPer............................................................... 552
TRIGger:EVENt<ev>:WINDow:DELTa.............................................................................. 552
TRIGger:EVENt<ev>:WINDow:RANGe............................................................................ 553
TRIGger:EVENt<ev>:WINDow:TIME................................................................................553
TRIGger:EVENt<ev>:WINDow:WIDTh............................................................................. 554
TRIGger:EVENt<ev>:LEVel<n>:WINDow:LOWer <Level>
Sets the lower voltage limit.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<Level> Range: -10 to 10
Increment: 0.001
*RST: -0.1
Default unit: V
Usage: Asynchronous command
Manual operation: See "Lower level" on page 156
TRIGger:EVENt<ev>:LEVel<n>:WINDow:UPPer <Level>
Sets the upper voltage limit.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<Level> Range: 0 to 1000
Increment: 1E-06
*RST: 0.1
Default unit: V
Usage: Asynchronous command
Manual operation: See "Upper level" on page 155
TRIGger:EVENt<ev>:WINDow:DELTa <WidthDelta>
Defines a range around the width value.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
Trigger