Remote control commandsMXO 4 Series
563User Manual 1335.5337.02 ─ 07
Parameters:
<Logic> AND | OR
*RST: AND
Usage: Asynchronous command
Manual operation: See "State Operator" on page 167
TRIGger:EVENt<ev>:STATe:QUALify:DIGital:CHAN<n>:HLX <HLX>
Sets the required state for each digital channel that is used for triggering.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 0 to 15
Parameters:
<HLX> HIGH | LOW | DONTcare
*RST: DONTcare
Usage: Asynchronous command
Manual operation: See "Source: channel states" on page 168
17.9.13 Pattern trigger
TRIGger:EVENt<ev>:PATTern:QUALify:ANALog:CHAN<n>:HLX........................................ 563
TRIGger:EVENt<ev>:PATTern:QUALify:DIGital:CHAN<n>:HLX.......................................... 564
TRIGger:EVENt<ev>:PATTern:QUALify:LOGic.................................................................. 564
TRIGger:EVENt<ev>:PATTern:QUALify:DIGital:LOGic....................................................... 564
TRIGger:EVENt<ev>:PATTern:QUALify:SOURces.............................................................564
TRIGger:EVENt<ev>:PATTern:QUALify:ANALog:CHAN<n>:HLX <HLX>
Set the state for each channel. For the state trigger, the clock source is indicated and
does not get a state.
Suffix:
<ev>
.
1 = A-trigger, 2 = B-trigger, 3 = reset event
<n> 1 to 4, index of the analog channel
Parameters:
<HLX> HIGH | LOW | DONTcare
State of the individual channels
*RST: DONTcare
Example:
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN1:HLX LOW
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN2:HLX HIGH
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN3:HLX LOW
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN4:HLX HIGH
TRIG:EVEN1:PATT:QUAL:LOG AND