Remote control commandsMXO 4 Series
580User Manual 1335.5337.02 ─ 07
Parameters:
<Stop> Range: -1E+26 to 1E+26
Increment: 0.01
*RST: 0.01
Usage: Asynchronous command
Manual operation: See "Start Stop" on page 179
LAYout<ly>:ZOOM<zo>:VERTical:MODE <Mode>
Defines if absolute or relative values are used to specify the y-axis values. Since the
zoom area refers to the active signal, relative values ensure that the zoom area
remains the same.
Suffix:
<ly>
.
1...4, index of the SmartGrid layout
<zo> 1...4, index of the zoom
Parameters:
<Mode> ABS | REL
*RST: REL
Usage: Asynchronous command
Manual operation: See "Mode" on page 180
LAYout<ly>:ZOOM<zo>:VERTical:RELative:POSition <RelativeCenter>
Defines the y-value of the centerpoint of the zoom area in relative values.
Suffix:
<ly>
.
1...4, index of the SmartGrid layout
<zo> 1...4, index of the zoom
Parameters:
<RelativeCenter> Range: 0 to 100
Increment: 0.1
*RST: 50
Default unit: %
Usage: Asynchronous command
Manual operation: See "Position range" on page 179
LAYout<ly>:ZOOM<zo>:VERTical:RELative:STARt <RelativeStart>
Defines the lower limit of the zoom area on the y-axis in relative values.
Suffix:
<ly>
.
1...4, index of the SmartGrid layout
<zo> 1...4, index of the zoom