Remote control commandsMXO 4 Series
596User Manual 1335.5337.02 ─ 07
REFCurve<rc>:RESCale:VERTical:OFFSet <Offset>
The vertical offset moves the reference waveform vertically. Enter a value with the unit
of the waveform.
Like vertical offset of a channel waveform, the offset of a reference waveform is sub-
tracted from the measured value. Negative values shift the waveform up, positive val-
ues shift it down.
Suffix:
<rc>
.
1...4, index of the reference waveform
Parameters:
<Offset> Range: -100000000 to 100000000
Increment: 1E-06
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "Vertical offset" on page 205
REFCurve<rc>:RESCale:VERTical:STATe <State>
If enabled, the vertical offset and factor are applied to the reference waveform.
Stretching and offset change the display of the waveform independent of the vertical
scale and position.
Suffix:
<rc>
.
1...4, index of the reference waveform
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Enable" on page 205
REFCurve<rc>:TOORiginal
Available, if REFCurve<rc>:VMODe is set to INDependent.
Restores the original vertical settings of the reference waveform (vertical scale, posi-
tion, and offset).
Suffix:
<rc>
.
1...4, index of the reference waveform
Usage: Setting only
Asynchronous command
Manual operation: See "Set to original" on page 203
Waveform analysis