Remote control commandsMXO 4 Series
617User Manual 1335.5337.02 ─ 07
Suffix:
<mg>
.
1...16, index of the measurement
Parameters:
<EdgesSlope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Edges slope" on page 233
MEASurement<mg>:AMPTime:PTCount
<PulseCount>
Sets the number of positive pulses for the pulse train measurement. It measures the
duration of N positive pulses from the rising edge of the first pulse to the falling edge of
the N-th pulse.
Suffix:
<mg>
.
1...16, index of the measurement
Parameters:
<PulseCount> Range: 1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Pulse count" on page 233
MEASurement<mg>:AMPTime:PSLope <PulsesSlope>
Sets the first slope of the pulses to be counted. The setting is only relevant for pulse
count measurement (MEASurement<mg>:MAIN is set to PULCnt.
Suffix:
<mg>
.
1...16, index of the measurement
Parameters:
<PulsesSlope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Pulse slope" on page 233
MEASurement<mg>:AMPTime:DELay<n>:DIRection <EdgeCountDir>
Selects the direction for counting slopes for each source: from the beginning of the
waveform, or from the end.
Suffix:
<mg>
.
1...16, index of the measurement
Automatic measurements