Remote control commandsMXO 4 Series
618User Manual 1335.5337.02 ─ 07
<n> 1..2
Selects the source number.
Parameters:
<EdgeCountDir> FRFI | FRLA
FRFI - FRom FIrst, counting starts with the first edge of the
waveform.
FRLA - FRom LAst, counting starts with the last edge of the
waveform.
*RST: FRFI
Usage: Asynchronous command
Manual operation: See "Direction" on page 235
MEASurement<mg>:AMPTime:DTOTrigger<n>:SLOPe
<DlyTrigSlp>
Sets the edge direction to be used for delay measurement.
Suffix:
<mg>
.
1...16, index of the measurement
<n> 1..2
Selects the source number.
Parameters:
<DlyTrigSlp> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Delay to trigger measurement settings" on page 235
MEASurement<mg>:AMPTime:DELay<n>:SLOPe <Slope>
Sets the edge of each source, between which the delay is measured.
Suffix:
<mg>
.
1...16, index of the measurement
<n> 1..2
Selects the source number.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
Usage: Asynchronous command
Manual operation: See "Slope" on page 235
Automatic measurements