17.12.6 Reference levels
REFLevel<rl>:ENABle.................................................................................................... 627
REFLevel<rl>:LMODe.................................................................................................... 627
REFLevel<rl>:ABSolute:HYSTeresis................................................................................ 627
REFLevel<rl>:ABSolute:LLEVel.......................................................................................628
REFLevel<rl>:ABSolute:MLEVel...................................................................................... 628
REFLevel<rl>:ABSolute:ULEVel...................................................................................... 628
REFLevel<rl>:RELative:HYSTeresis.................................................................................629
REFLevel<rl>:RELative:LOWer....................................................................................... 629
REFLevel<rl>:RELative:MIDDle.......................................................................................629
REFLevel<rl>:RELative:MODE........................................................................................630
REFLevel<rl>:RELative:UPPer........................................................................................630
REFLevel<rl>:ENABle <First>
Enables the specified reference level.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<First> ON | OFF
Usage: Asynchronous command
Manual operation: See "RLx" on page 224
REFLevel<rl>:LMODe <LevelMode>
Defines if the reference level is set in absolute or relative values.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<LevelMode> ABS | REL
*RST: REL
Usage: Asynchronous command
Manual operation: See "Level mode" on page 225
REFLevel<rl>:ABSolute:HYSTeresis <HystAbs>
Defines a hysteresis for the middle reference level. A rise or fall from the middle refer-
ence value that does not exceed the hysteresis is rejected as noise.
Suffix:
<rl>
.
1...4, index of the reference level set
Automatic measurements