EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Reference Levels

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
627User Manual 1335.5337.02 ─ 07
17.12.6 Reference levels
REFLevel<rl>:ENABle.................................................................................................... 627
REFLevel<rl>:LMODe.................................................................................................... 627
REFLevel<rl>:ABSolute:HYSTeresis................................................................................ 627
REFLevel<rl>:ABSolute:LLEVel.......................................................................................628
REFLevel<rl>:ABSolute:MLEVel...................................................................................... 628
REFLevel<rl>:ABSolute:ULEVel...................................................................................... 628
REFLevel<rl>:RELative:HYSTeresis.................................................................................629
REFLevel<rl>:RELative:LOWer....................................................................................... 629
REFLevel<rl>:RELative:MIDDle.......................................................................................629
REFLevel<rl>:RELative:MODE........................................................................................630
REFLevel<rl>:RELative:UPPer........................................................................................630
REFLevel<rl>:ENABle <First>
Enables the specified reference level.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<First> ON | OFF
Usage: Asynchronous command
Manual operation: See "RLx" on page 224
REFLevel<rl>:LMODe <LevelMode>
Defines if the reference level is set in absolute or relative values.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<LevelMode> ABS | REL
*RST: REL
Usage: Asynchronous command
Manual operation: See "Level mode" on page 225
REFLevel<rl>:ABSolute:HYSTeresis <HystAbs>
Defines a hysteresis for the middle reference level. A rise or fall from the middle refer-
ence value that does not exceed the hysteresis is rejected as noise.
Suffix:
<rl>
.
1...4, index of the reference level set
Automatic measurements