Remote control commandsMXO 4 Series
628User Manual 1335.5337.02 ─ 07
Parameters:
<HystAbs> Range: 0 to 1E+26
Increment: 0.001
*RST: 0.005
Default unit: V
Usage: Asynchronous command
Manual operation: See "Hysteresis" on page 225
REFLevel<rl>:ABSolute:LLEVel <LowerLevel>
Sets the lower reference level in absolute values. This is required, e.g., to determine a
fall.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<LowerLevel> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "Upper level, Middle level, Lower level" on page 225
REFLevel<rl>:ABSolute:MLEVel <MiddleLevel>
Sets the middle reference level in absolute values.
Suffix:
<rl>
.
1...4, index of the reference level set
Parameters:
<MiddleLevel> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "Upper level, Middle level, Lower level" on page 225
REFLevel<rl>:ABSolute:ULEVel <UpperLevel>
Sets the upper reference level in absolute values. This is required to determine a rise.
Suffix:
<rl>
.
1...4, index of the reference level set