Remote control commandsMXO 4 Series
642User Manual 1335.5337.02 ─ 07
Return values:
<Delta> Range: -1E+26 to 1E+26
Increment: 0.1
*RST: 0
Default unit: s
Usage: Query only
Asynchronous command
CURSor<cu>:XDELta:INVerse?
Queries the inverse value of the delta value (distance) of the two vertical cursor lines.
Suffix:
<cu>
.
1...2, index of the cursor set
Return values:
<DeltaInverse> Range: -1E+26 to 1E+26
Increment: 0.1
*RST: 0
Default unit: Hz
Usage: Query only
Asynchronous command
CURSor<cu>:YDELta[:VALue]?
Queries the delta value (distance) of the two horizontal cursor lines.
Suffix:
<cu>
.
1...2, index of the cursor set
Return values:
<Delta> Range: -1E+26 to 1E+26
Increment: 0
*RST: 0
Usage: Query only
Asynchronous command
CURSor<cu>:YDELta:SLOPe <DeltaSlope>
Returns the inverse value of the voltage difference - the reciprocal of the vertical dis-
tance of two horizontal cursor lines: 1/
Δ
V.
Suffix:
<cu>
.
1...2, index of the cursor set
Parameters:
<DeltaSlope> Range: -1E+26 to 1E+26
Increment: 0
*RST: 0
Cursor measurements