Remote control commandsMXO 4 Series
652User Manual 1335.5337.02 ─ 07
CALCulate:SPECtrum<sp>:PEXCursion <PeakExcursion>
Defines a minimum level value by which the waveform must drop left and right of the
local maximum to be listed as a peak. Enter a peak excursion value to omit close by
peaks and list just the highest peak.
Suffix:
<sp>
.
1
Parameters:
<PeakExcursion> Range: 0 to 100
Increment: 1
*RST: 5
Default unit: dB
Usage: Asynchronous command
Manual operation: See "Peak excursion" on page 216
CALCulate:SPECtrum<sp>:WAVeform:AVERage:COUNt
<AverageCount>
Sets the number of segments used for the averaging of the spectrum.
Suffix:
<sp>
.
1
Parameters:
<AverageCount> Range: 2 to 65534
Increment: 1
*RST: 1000
Usage: Asynchronous command
Manual operation: See "Traces" on page 249
CALCulate:SPECtrum<sp>:WAVeform:AVERage:ENABle <Enable>
Enables the average trace.
Suffix:
<sp>
.
1
Parameters:
<Enable> ON | OFF
Usage: Asynchronous command
Manual operation: See "Traces" on page 249
CALCulate:SPECtrum<sp>:WAVeform:MAXimum:ENABle <Enable>
Enables the maximum trace.
Suffix: