Remote control commandsMXO 4 Series
657User Manual 1335.5337.02 ─ 07
Manual operation: See "Peak on waveform" on page 253
CALCulate:SPECtrum<sp>:PLISt:RESult[:VALue]?
Returns the current peak list measurement results.
Suffix:
<sp>
.
1
Usage: Query only
Asynchronous command
CALCulate:SPECtrum<sp>:PLISt:SORT:COLumn
<ResultColumn>
Sorts the results in the spectrum peak list table either according to the frequency or
according to the value.
Suffix:
<sp>
.
1
Parameters:
<ResultColumn> FREQ | VAL
*RST: FREQ
Usage: Asynchronous command
CALCulate:SPECtrum<sp>:PLISt:SORT:ORDer <ResultOrder>
Defines if the spectrum peak list results are sorted in an ascending (increasing) or
descending (decreasing) order.
Suffix:
<sp>
.
1
Parameters:
<ResultOrder> ASC | DESC
ASC: ascending
DESC: descending
*RST: ASC
Usage: Asynchronous command
CALCulate:SPECtrum<sp>:PLISt:LABel:FREQuency[:STATe] <ShowFrequency>
Includes the frequency of the detected peak in the diagram labels.
Suffix:
<sp>
.
1
Parameters:
<ShowFrequency> ON | OFF
*RST: ON
Spectrum analysis