Remote control commandsMXO 4 Series
665User Manual 1335.5337.02 ─ 07
FRANalysis:POINts:MODE <PointsMode>
Selects, if the number of points for the FRA are measured as total or per decade.
You can set the number of points with FRANalysis:POINts:TOTal/ FRANalysis:
POINts:LOGarithmic.
Parameters:
<PointsMode> TOTal | DECade
*RST: DECade
Usage: Asynchronous command
Manual operation: See "Points" on page 261
FRANalysis:POINts:TOTal <TotalPoints>
Set the total number of points for the FRA analysis, if FRANalysis:POINts:MODE
on page 665 is set to TOTal.
Parameters:
<TotalPoints> Range: 10 to 5000
Increment: 1
*RST: 100
Usage: Asynchronous command
Manual operation: See "Points" on page 261
FRANalysis:REPeat <Repeat>
Repeats the measurement, using the same parameters.
Parameters:
<Repeat> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Repeat" on page 261
FRANalysis:RESet
Resets the frequency response analysis.
Usage:
Setting only
Asynchronous command
FRANalysis:PHASe:MAXimum <MaxPhase>
Sets the upper boundary of the vertical phase window.
The lower boundary is given by "Maximum phase" - 360°.