FRANalysis:AMPLitude:PROFile:SORT
Sorts the points in the amplitude table by frequency, starting with the lowest frequency.
Usage: Setting only
Asynchronous command
Manual operation: See "Sort" on page 263
FRANalysis:AMPLitude:PROFile:MODE <VoltageChange>
Selects if the voltage change is done as a single step or as a ramp.
Parameters:
<VoltageChange> SINGle | RAMP
*RST: SINGle
Usage: Asynchronous command
Manual operation: See "Voltage change" on page 262
FRANalysis:AMPLitude:PROFile:POINt<m>:FREQuency <Frequency>
Sets the start frequency for the selected point.
Suffix:
<m>
.
Index of the amplitude profile point
Parameters:
<Frequency> Range: 0.01 to 100E6
Increment: 1
*RST: 100
Default unit: Hz
Usage: Asynchronous command
Manual operation: See "Step start freq, Amplitude" on page 262
FRANalysis:AMPLitude:PROFile:POINt<m>:AMPLitude <Amplitude>
Sets the amplitude for the selected point.
Suffix:
<m>
.
Index of the amplitude profile point
Parameters:
<Amplitude> Range: 0 to 5.995
Increment: 0.1
*RST: 1
Default unit: Vpp
Usage: Asynchronous command
Manual operation: See "Step start freq, Amplitude" on page 262